The natural distribution of stress in the femur is altered when total hip arthroplasty (THA) is performed. In fact, when a stem is inserted inside the femur, there is a variation in stress due to the difference in rigidity between the material with which the stem is made and the femur. This generates the phenomenon of stress shielding. The aim of this study is to design an optimized prosthesis that guarantees an excellent rotational stability and a reduced stress shielding. Methods: Through the finite element method (FEM), the mechanical behavior of the stem subjected to the loads described by ISO 7206-4:2010 is studied. Results: Through topological optimization, there is a reduction in stress shielding in the proximal zone of 31.46%. The addition of ridges on the dorsal side of the stem also improves rotational stability by 27.82%. Conclusions: The decrease in stiffness that is recorded with the optimized stem guarantees a greater distribution of stress on the bone. The presence of dorsal ridges also favors the corticalization of the bone as it loads the bone near the dorsal, ensuring further stability. The perforated prosthesis presented in this study shows an increase in primary stability and an improvement in rotational stability as there is also a bone regrowth inside the prosthesis.

Evaluation of Rotational Stability and Stress Shielding of a Stem Optimized for Hip Replacements-A Finite Element Study / Ceddia, M; Trentadue, B. - In: PROSTHESIS. - ISSN 2673-1592. - 5:3(2023), pp. 678-693. [10.3390/prosthesis5030048]

Evaluation of Rotational Stability and Stress Shielding of a Stem Optimized for Hip Replacements-A Finite Element Study

Ceddia, M
Membro del Collaboration Group
;
Trentadue, B
Writing – Review & Editing
2023-01-01

Abstract

The natural distribution of stress in the femur is altered when total hip arthroplasty (THA) is performed. In fact, when a stem is inserted inside the femur, there is a variation in stress due to the difference in rigidity between the material with which the stem is made and the femur. This generates the phenomenon of stress shielding. The aim of this study is to design an optimized prosthesis that guarantees an excellent rotational stability and a reduced stress shielding. Methods: Through the finite element method (FEM), the mechanical behavior of the stem subjected to the loads described by ISO 7206-4:2010 is studied. Results: Through topological optimization, there is a reduction in stress shielding in the proximal zone of 31.46%. The addition of ridges on the dorsal side of the stem also improves rotational stability by 27.82%. Conclusions: The decrease in stiffness that is recorded with the optimized stem guarantees a greater distribution of stress on the bone. The presence of dorsal ridges also favors the corticalization of the bone as it loads the bone near the dorsal, ensuring further stability. The perforated prosthesis presented in this study shows an increase in primary stability and an improvement in rotational stability as there is also a bone regrowth inside the prosthesis.
2023
Evaluation of Rotational Stability and Stress Shielding of a Stem Optimized for Hip Replacements-A Finite Element Study / Ceddia, M; Trentadue, B. - In: PROSTHESIS. - ISSN 2673-1592. - 5:3(2023), pp. 678-693. [10.3390/prosthesis5030048]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11589/262180
Citazioni
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact