Laser scanning thermography (LST) is an effective nondestructive testing (NDT) method for detecting defects in carbon fiber reinforced polymer (CFRP). The thermal camera is the most critical measurement unit in LST defect detection technology, yet research gaps remain regarding the impact of its thermal sensitivity and spatial resolution on defect detection. To address the issue, this article proposes a systematic evaluation method of the effects of noise equivalent temperature difference (NETD) and spatial resolution on CFRP internal defect detection using joint LST (JLST). Based on heat conduction theory, the relationship between NETD and detectable defect depth is analytically derived, establishing a measurement sensitivity limit. The effect of spatial resolution on detectable defect size is further analyzed from a sampling perspective. Moreover, a defect signal-to-noise ratio (SNR) model is proposed to characterize the coupled effects of NETD and spatial resolution across different diameter-to-depth ratios (DDRs). The experiments are carried out on the CFRP panel with flat bottom hole defects using two cameras with different NETD and spatial resolution. The effects of these two parameters were highlighted by comparing the results with other thermography defect detection methods. The results demonstrate that NETD contributes more to detecting small DDR defects at the same diameter, while spatial resolution dominates for small DDR defects at the same depth. This study provides practical guidance for the specifications of thermographic cameras and the improvement of defect detection accuracy in the laser thermography-based detection of internal defects in CFRP.

Quantitative Evaluation of Thermal Camera Thermal Sensitivity and Spatial Resolution on Internal Defect Detection in CFRP Using Laser Thermography / He, Y., Wu, H., Deng, B., Yang, R., Yu, Y., Ciampa, F.. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 75:(2026), pp. 3516912.3516912-3516912.3516912. [10.1109/tim.2026.3704085]

Quantitative Evaluation of Thermal Camera Thermal Sensitivity and Spatial Resolution on Internal Defect Detection in CFRP Using Laser Thermography

Ciampa, Francesco
2026

Abstract

Laser scanning thermography (LST) is an effective nondestructive testing (NDT) method for detecting defects in carbon fiber reinforced polymer (CFRP). The thermal camera is the most critical measurement unit in LST defect detection technology, yet research gaps remain regarding the impact of its thermal sensitivity and spatial resolution on defect detection. To address the issue, this article proposes a systematic evaluation method of the effects of noise equivalent temperature difference (NETD) and spatial resolution on CFRP internal defect detection using joint LST (JLST). Based on heat conduction theory, the relationship between NETD and detectable defect depth is analytically derived, establishing a measurement sensitivity limit. The effect of spatial resolution on detectable defect size is further analyzed from a sampling perspective. Moreover, a defect signal-to-noise ratio (SNR) model is proposed to characterize the coupled effects of NETD and spatial resolution across different diameter-to-depth ratios (DDRs). The experiments are carried out on the CFRP panel with flat bottom hole defects using two cameras with different NETD and spatial resolution. The effects of these two parameters were highlighted by comparing the results with other thermography defect detection methods. The results demonstrate that NETD contributes more to detecting small DDR defects at the same diameter, while spatial resolution dominates for small DDR defects at the same depth. This study provides practical guidance for the specifications of thermographic cameras and the improvement of defect detection accuracy in the laser thermography-based detection of internal defects in CFRP.
2026
Quantitative Evaluation of Thermal Camera Thermal Sensitivity and Spatial Resolution on Internal Defect Detection in CFRP Using Laser Thermography / He, Y., Wu, H., Deng, B., Yang, R., Yu, Y., Ciampa, F.. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 75:(2026), pp. 3516912.3516912-3516912.3516912. [10.1109/tim.2026.3704085]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11589/310160
Citazioni
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 10
  • OpenAlex 15
social impact