The aim of this paper was to investigate the diagnostic potential of tomosynthesis imaging compared with the performance of 2-D digital mammography in terms of radiation dose and image quality. In particular, suitable dosimeter and phantom were used for quantifying the average glandular dose and image quality parameters, respectively. First, according to standard protocols and European guidelines, the characterization of the used tomosynthesis system was carried out to verify the reliability of characteristic parameters of the system. Successively, the absorbed dose was calculated by means of experimental measurements and the application of estimation methods. The calculated dose was then compared with the value provided by the system; this approach has confirmed the tendency of mammography equipment manufacturers to underestimate the mean glandular dose. Finally, the detection capability of different details with different contrasts was objectively assessed for both breast tomosynthesis and 2-D mammography.
Dosimetric characterization and image quality assessment in Breast Tomosynthesis / Andria, Gregorio; Attivissimo, Filippo; DI NISIO, Attilio; Lanzolla, Anna Maria Lucia; Maiorana, A; Mangiantini, M; Spadavecchia, Maurizio. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - STAMPA. - 66:10(2017), pp. 2535-2544. [10.1109/TIM.2017.2692318]
Dosimetric characterization and image quality assessment in Breast Tomosynthesis
ANDRIA, Gregorio;ATTIVISSIMO, Filippo;DI NISIO, Attilio;LANZOLLA, Anna Maria Lucia;SPADAVECCHIA, MAURIZIO
2017-01-01
Abstract
The aim of this paper was to investigate the diagnostic potential of tomosynthesis imaging compared with the performance of 2-D digital mammography in terms of radiation dose and image quality. In particular, suitable dosimeter and phantom were used for quantifying the average glandular dose and image quality parameters, respectively. First, according to standard protocols and European guidelines, the characterization of the used tomosynthesis system was carried out to verify the reliability of characteristic parameters of the system. Successively, the absorbed dose was calculated by means of experimental measurements and the application of estimation methods. The calculated dose was then compared with the value provided by the system; this approach has confirmed the tendency of mammography equipment manufacturers to underestimate the mean glandular dose. Finally, the detection capability of different details with different contrasts was objectively assessed for both breast tomosynthesis and 2-D mammography.File | Dimensione | Formato | |
---|---|---|---|
Post print 2017 TIM Dosimetric Characterization and Image Quality Assessment in Breast Tomosynthesis-convertito.pdf
accesso aperto
Descrizione: Accepted version
Tipologia:
Documento in Post-print
Licenza:
Tutti i diritti riservati
Dimensione
1.22 MB
Formato
Adobe PDF
|
1.22 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.