The main purpose of this paper is to analyze the digital radiographic methodologies used in diagnostic investigation, in order to reduce radiation doses to the patients by assuring a good quality of the images. To this end, the main parameters influencing the absorbed dose during radiographic exams and the effect of the human body on X-ray attenuation were investigated. It is generally understood that radiographic techniques need an appropriate choice of the operative parameters as functions of patients' characteristics, with the aim of providing the minimum dose that is compatible with an image that satisfies the goal of the examination and has a good signal-to-noise ratio. In an effort to identify an optimized radiographic protocol, the correlation between body size and radiation dose was investigated.

Dose Optimization in Chest Radiography: System and Model Characterization via Experimental Investigation / Andria, Gregorio; Attivissimo, Filippo; DI NISIO, Attilio; Lanzolla, Anna Maria Lucia; G., Guglielmi; R., Terlizzi. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 63:5(2014), pp. 1163-1170. [10.1109/TIM.2013.2282411]

Dose Optimization in Chest Radiography: System and Model Characterization via Experimental Investigation

ANDRIA, Gregorio;ATTIVISSIMO, Filippo;DI NISIO, Attilio;LANZOLLA, Anna Maria Lucia;
2014-01-01

Abstract

The main purpose of this paper is to analyze the digital radiographic methodologies used in diagnostic investigation, in order to reduce radiation doses to the patients by assuring a good quality of the images. To this end, the main parameters influencing the absorbed dose during radiographic exams and the effect of the human body on X-ray attenuation were investigated. It is generally understood that radiographic techniques need an appropriate choice of the operative parameters as functions of patients' characteristics, with the aim of providing the minimum dose that is compatible with an image that satisfies the goal of the examination and has a good signal-to-noise ratio. In an effort to identify an optimized radiographic protocol, the correlation between body size and radiation dose was investigated.
2014
Dose Optimization in Chest Radiography: System and Model Characterization via Experimental Investigation / Andria, Gregorio; Attivissimo, Filippo; DI NISIO, Attilio; Lanzolla, Anna Maria Lucia; G., Guglielmi; R., Terlizzi. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 63:5(2014), pp. 1163-1170. [10.1109/TIM.2013.2282411]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11589/1088
Citazioni
  • Scopus 26
  • ???jsp.display-item.citation.isi??? 21
social impact