In this paper a solution for metrological calibration of digital instruments without built-in communication interface (DIWIs) is presented. The solution is based on the conversion in numerical data of video camera images of DIWI display. The paper describes the vision algorithm used, illustrates the functionality of the realized software (in LabVIEW environment), and presents the results of an actual calibration experiment performed on a Fluke model 189 True RMS digital multimeter.
Automatic calibration system for digital instruments without built-in communication interface / Andria, Gregorio; Cavone, Giuseppe; Fabbiano, L.; Giaquinto, Nicola; Savino, Mario. - 3:(2009), pp. 1934-1937. (Intervento presentato al convegno 19th IMEKO World Congress 2009 tenutosi a Lisbon, Portugal nel 6-11 September 2009).
Automatic calibration system for digital instruments without built-in communication interface
ANDRIA, Gregorio;CAVONE, Giuseppe;Fabbiano, L.;GIAQUINTO, Nicola;SAVINO, Mario
2009-01-01
Abstract
In this paper a solution for metrological calibration of digital instruments without built-in communication interface (DIWIs) is presented. The solution is based on the conversion in numerical data of video camera images of DIWI display. The paper describes the vision algorithm used, illustrates the functionality of the realized software (in LabVIEW environment), and presents the results of an actual calibration experiment performed on a Fluke model 189 True RMS digital multimeter.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.